Key features
- 16-channel system allowing for 8 or 16 simultaneous reads
- 96 and 384-well format
- measure up to 816 samples per run
- Featuring sample temperature and evaporation control
- Minimal sample volume of 40 µL
- 21 CFR Part 11 compliant software available
Sales in
Austria, Belgium, Germany, Luxembourg, Switzerland, The Netherlands
Gator Pivot – Biolayer Interferometry (BLI) system
Gator® systems utilise a new generation of label-free analysis technology integrated into an innovative biolayer interferometry (BLI) instrument platform. The advanced BLI technology detects biomolecular interactions with high sensitivity, accuracy and speed. The Gator® Pivot BLI instrument is the new BLI system from Gator Bio, designed for thermally unstable proteins and experiments requiring extended assay times. The instrument has 16 spectrophotometers and a flexible 2-plate design (96- and 384-wells), allowing quantitation of 16 samples simultaneously with a maximum of 816 samples per run.
The Gator Pivot has an incorporated evaporation control enabling sample temperature regulation between 15-40⁰ Celcius and extended assay times. The instrument is ready for automated plate handling and features automated cleaning protocols. The incorporated software facilitates easy assay setup, seamless data acquisition, and streamlined analysis. Additionally, Gator Bio delivers a cutting-edge portfolio of biosensors featuring improved surface chemistry to boost sensitivity, binding capacity, robustness, and regenerative capabilities.
Gator Bio BLI systems provide a platform for analysing proteins, antibodies, peptides, nucleic acids, liposomes, viruses and small molecules.
Applications are (but not limited to):
-
- quantitation and kinetics
- epitope binning
- bioprocess analytics and manufacturing control
- ELISA replacement
- viruses and vaccine development
- antibody discovery
Downloads
Gator family brochure
Gator family one pager
Gator family spec sheet
Introduction brochure
Technical data
Type of Analysis | Proteins, antibodies, peptides, nucleic acids, liposomes, viruses, small molecules |
Simultanious Reads | 8 or 16 |
Maximum Sample Capacity | 816 |
Molecular Weight | > 150 Da |
Association Rate (kon) | 101 to 107 M-1 s-1 |
Dissociation Rate (koff) | 10-6- 10-1 s-1 |
Affinity Constant (KD) | 10 pM – 1 mM |
Quantitation Range (Protein A biosensor) | 0.02 – 2000 µg/mL |
Binning Capacity | 20×20 |
Baseline Noise (RMS) | ≤ 4 pm |
Baseline Drift | ≤ 0.1 nm/hour |
Acquisition Rate | 2, 5, and 10 Hz |
Specifications | |
Spectrometers | 16 |
Sample Microplate | 2×96 or 384-well format |
Evaporation Control | Yes |
Sample temperature Control | 15°C to 40°C |
Automation compatible | Yes |
Minimum sample volume | 40 µL |
Smart monitoring | Yes |
Self-cleaning | YES |
Dimension – HxWxD (cm) | 92x87x79 |
Weight (kg) | 130 kg |