Key features

  • Pre-calibrated, GMP-ready temperature sensors with unique ID storing calibration data
  • Accurate monitoring to maintain validated temperature ranges in critical processes
  • Helps prevent protein denaturation caused by temperature rises during filtration and pump action
  • Five connection sizes with full NIST traceability, scalable from development to GMP production

Sales in

Austria, Belgium, Germany, Luxembourg, Switzerland, The Netherlands

Reliable Temperature Monitoring for Critical Bioprocesses

Temperature control is essential in biopharmaceutical production, where many filtration and separation steps are validated within strict ranges. SciLog® SciTemp® sensors ensure these validated conditions are met, supporting reliable virus filtration and other sensitive applications.

By providing precise monitoring, the sensors help prevent protein denaturation caused by temperature rises during processes such as pumping or filtration. This safeguards product integrity and maintains process consistency.

Each pre-calibrated, GMP-suitable sensor carries a unique ID with stored calibration data. With five available connection sizes and full NIST traceability, SciTemp® sensors offer scalability from development to full GMP production environments.

SciTemp® Temperature Sensor Monitor

The SciTemp® Monitor provides precise, real-time temperature measurement for SciLog® Single-Use Temperature Sensors. It supports up to two sensor inputs simultaneously and displays T1, T2 and differential temperature (DT). Within the 4.00° to 70.00° C range, the monitor delivers an accuracy of ±0.5° C.
For added process control, the SciTemp® Monitor includes three TTL alarm switches with user-selectable high and low limits, and offers both analogue and digital (RS-232) outputs for seamless integration into existing data systems.

Downloads

SciTemp Sensors Datasheet

Technical data

Sensors
Industry Life Sciences
Brand Parker Hannifin
Application Temperature
Wetted Material Animal-Free Compounds
Materials of Construction Medical Grade Polysulfone
Specifications Met USP Class VI, FDA 21 CRF 177.1520
Performance Temperature Range: 4 to 70°C
Monitor
Height 2.75 inch, 69.85 mm
Width 6 inch, 152.4 mm
Depth: 5.75 inch, 146.05 mm
Weight: 1.30073 lb, 0.59 kg