Key features
- Photometric measurement
- High sample capacity, up to 126 samples at once with continuous loading
- Measurement of up to 28 parameter in parallel (2-40 µl sample/parameter)
- Monitoring of mammalian, bacterial and yeast culture systems
- Automated daily maintenance
- Freely programmable automated on-board sample dilution
Sales in
Austria, Belgium, Denmark, France, Germany, Ireland, Luxembourg, Sweden, Switzerland, The Netherlands, United Kingdom
Cedex® Bio HT – Bioanalyzer for bioprocess control
The Cedex® Bio HT Analyzer is a multi-analyser for monitoring bioprocesses and fermentations. The wide range of measurable parameters allows the Cedex® Bio HT Analyzer to monitor the production and to develop and optimise growth conditions for mammalian, bacteria, and yeast culture systems. Substrates, metabolites, and products are measured photometrically by the Cedex Bio HT Analyzer, which is recommended for high throughput screening. The Cedex Bio HT Analyzer processes numerous samples simultaneously with minimal manual intervention.
The 96-well upgrade kit allows for the use of 96-well plates for a total capacity of up to 96 + 30 samples. Automated sample dilution saves time and ensures high accuracy. For process optimisation, the full automation of the Cedex® Bio HT also allows online integration with other automated systems.
Are you interested in automating the data handling of your Cedex Bio HT Analyzer?
Then, look at Sm@rtLine Data Cockpit (SDC) from our partner AGU. SDC is a middleware for central data handling in production, development, and research. It enables the connection of the Cedex Bio HT Analyzer (and many more analysers) to your LIMS system.
Contact us to figure out if SDC is the solution to your requirements.
Downloads
Brochure
SDC Brochure
Cedex® Bio HT Analyzer Multiwell Upgrade
Technical data
Method of Measurement | Photometric |
Test Throughput (Theoretical Max) | 200-400 tests/hour |
Reagent capacity | 32 slots (approximatly 28 assays) |
Programmable Parameters | yes |
Sample Types | mammalian cell culture microbial fermentation yeast cell cultivation |
Sample Input | Up to 90 sample convenience tubes Priority sampling available Barcode reading 96 well-plate upgrade kit (96 + 30 samples at once) |
Sample Volume | 2-40 μL sample consumption per test |
Minimum Sample Volume | 100 μL for several tests |
Cuvette capacity | 1000 |
pre-dilution | free selectable |
Rerun (automatic dilution) | re-run with automatic dilutions |
Sample Barcode Types | Code 128, Codabar, Code 2 of 5 interleaved, code 39 |
Sample cooling | yes |
System Interfaces | RS232 serial interface for HOST connection, Ethernet for remote access Online integration with automated systems |
Operating Conditions | Ambient temperature: 15 – 32°C Ambient humidity: 30 – 80% (RH, without condensation) Water purity: Reagent grad NCCLS Typ I Water consumption: 2 L/day – connection to water supply possible (optional) Waste: Separate container – connection to sink possible |
Control Unit | HP workstation running Windows 10, Intel Core i5 with 8 GB RAM Dual 500 GB hard drives, DVDRW drive, Ethernet |
Electrical Requirements | Power requirements: 100 – 125 V and 200 – 240 V Consumption: 1200 VA Frequency: 50 Hz oder 60 Hz |
Physical Dimensions | Width: 1350 mm Depth: 660 mm Height: 750 mm Weight: 198 kg (436 lbs) |
Certificates | CE, UL, C-UL |